Comparison of Transbronchial Lung Biopsy Yield between Standard Forceps and Electrocautery Hot Forceps in Swine

Wahidi M.M.a · Shofer S.L.a · Sporn T.A.b · Ernst A.c
aDivision of Pulmonary and Critical Care Medicine, Department of Internal Medicine, and bDepartment of Pathology, Duke University Medical Center, Durham, N.C., and cDivision of Pulmonary and Critical Care Medicine, Department of Internal Medicine, Beth Israel Deaconess Medical Center, Boston, Mass., USA
email Corresponding Author

Abstract

Background: Transbronchial lung biopsy (TBLB) is a commonly performed bronchoscopic procedure. Previous studies have suggested that larger biopsy forceps may improve diagnostic yield; however, the risk of bleeding associated with larger samples may be increased. The hot forceps are large forceps that are connected to an electrocautery system to minimize bleeding at the time of biopsy. Objectives: We evaluated the hot forceps for improvement in biopsy size and the number of sampled alveoli. Methods: TBLBs were performed in 2 swine using one type of the forceps, followed by the other forceps 24 h later. Electrocautery was applied from closure of the forceps to retrieval of the sample. A blinded pathologist measured the size of each sample in its longest dimension and calculated the total alveolar content within the largest cross-section from each biopsy. Results: A total of 74 biopsies were collected using each forceps type. Alveolar tissue was present in 25/74 and 26/74 of the biopsies using the hot and conventional forceps, respectively. There was no difference in the size of biopsies collected (2.10 ± 1.10 vs. 1.83 ± 0.94 mm; p = 0.164) or in the amount of alveoli per sample (343.2 ± 402.4 vs. 439.5 ± 463.5 alveoli; p = 0.433) for hot and conventional forceps, respectively. There was no artifact related to the use of electrocautery, and bleeding was minimal using either forceps system. Conclusions: The use of the electrocautery hot forceps for TBLB did not result in improvement of the size of biopsies or the amount of collected alveolar tissue in healthy pigs.


 Outline


 goto top of outline Key Words

  • Bronchoscopy
  • Transbronchial lung biopsy
  • Electrocautery

 goto top of outline Abstract

Background: Transbronchial lung biopsy (TBLB) is a commonly performed bronchoscopic procedure. Previous studies have suggested that larger biopsy forceps may improve diagnostic yield; however, the risk of bleeding associated with larger samples may be increased. The hot forceps are large forceps that are connected to an electrocautery system to minimize bleeding at the time of biopsy. Objectives: We evaluated the hot forceps for improvement in biopsy size and the number of sampled alveoli. Methods: TBLBs were performed in 2 swine using one type of the forceps, followed by the other forceps 24 h later. Electrocautery was applied from closure of the forceps to retrieval of the sample. A blinded pathologist measured the size of each sample in its longest dimension and calculated the total alveolar content within the largest cross-section from each biopsy. Results: A total of 74 biopsies were collected using each forceps type. Alveolar tissue was present in 25/74 and 26/74 of the biopsies using the hot and conventional forceps, respectively. There was no difference in the size of biopsies collected (2.10 ± 1.10 vs. 1.83 ± 0.94 mm; p = 0.164) or in the amount of alveoli per sample (343.2 ± 402.4 vs. 439.5 ± 463.5 alveoli; p = 0.433) for hot and conventional forceps, respectively. There was no artifact related to the use of electrocautery, and bleeding was minimal using either forceps system. Conclusions: The use of the electrocautery hot forceps for TBLB did not result in improvement of the size of biopsies or the amount of collected alveolar tissue in healthy pigs.

Copyright © 2009 S. Karger AG, Basel


goto top of outline Introduction

Flexible bronchoscopy with transbronchial lung biopsy (TBLB) is a commonly used technique in pulmonary medicine for the diagnosis of a wide variety of diseases. However, one of the limiting features of this technique is the small sample obtained using conventional forceps [1]. Previous studies have suggested that larger forceps may improve diagnostic yield; however, as forceps become larger, there is greater concern about the development of complications, particularly bleeding [2,3,4]. In recent years, larger forceps connected to an electrocautery system, named the ‘hot forceps’, were designed for use with flexible bronchoscopy. With these forceps, heat energy is delivered after grasping the tissue to promote coagulation and reduce bleeding at the time of tissue collection. A recent study evaluated these hot forceps for biopsies of endobronchial lesions and showed no difference in diagnostic yield and a small improvement in bleeding compared with conventional forceps [5]. No data are available regarding the use of these forceps for TBLBs. The objective of our study is to evaluate the hot forceps for improvement in the size of biopsies or the amount of collected alveolar tissue per sample.

 

goto top of outline Methods

goto top of outline Study Design

This investigation was designed as a cross-over blinded study. The pathologist assessing the histological biopsy samples was blinded to the type of forceps used during the experiment. The main outcome was the quantity of lung tissue, macroscopically and microscopically, obtained with TBLB.

goto top of outline Animals

The experiment protocol was approved by the Duke University Institutional Animal Care and Use Committee. The experiment was conducted in accordance with the National Research Council guidelines. Yorkshire swine were used for the study due to similarity of lung anatomy and mucosal and tissue characteristics to those of humans [6]. Additionally, pig models have been used previously in bronchoscopy research on airway stenosis, malacia, airway stents and TBLB [7,8,9,10].

Animals were procured for medical purposes and were free of lung diseases. They were housed in the animal facility at Duke University and were quarantined for 48 h prior to study entry.

goto top of outline Description of Procedure

The procedure was performed after an overnight fast. Swine were sedated with acepromazine 1.1 mg/kg intramuscularly and ketamine 22 mg/kg intramuscularly followed by insertion of an intravenous catheter into an ear vein. Propofol 6 mg/kg intravenously was then administered to achieve deeper sedation. Animals were endotracheally intubated and ventilated using a Respironics Esprit ventilator (Respironics, Murraysville, Pa., USA) set to deliver a tidal volume of 8 ml/kg and an O2 saturation of >95%. Blood pressure, O2 saturation and electrocardiogram tracing were monitored continuously throughout the procedure. The swine were sacrificed after the biopsies were completed.

Two animals, of equal size and weight, were enrolled in the study. One animal underwent TBLB with the standard forceps followed by the hot forceps 24 h later, while the other animal underwent the procedure in the opposite order.

All procedures were performed using a Pentax bronchoscope (Pentax, Inc., Montvale, N.J., USA) dedicated for animal use.

The conventional forceps cups (Conmed Corporation, Utica, N.Y., USA) were 1.8 mm wide (diameter A) and 2 mm long (diameter C), with a total cup capacity of 3 mm3 (fig. 1, 2). The hot forceps cups (Olympus, Inc., Center Valley, Pa., USA) were 2.2 mm wide (diameter A) and 2.6 mm long (diameter C), with a total cup capacity of 5.1 mm3.

FIG01
Fig. 1. Various dimensions (diameters A–C) for forceps cups.

FIG02
Fig. 2. Conventional (left) and hot electrocautery forceps (right).

The bronchoscope was introduced via the endotracheal tube. The right caudal lobe was identified by visual inspection and the tip of the bronchoscope was placed in the wedge position. TBLBs were performed under fluoroscopic guidance using conventional or hot forceps. The conventional forceps were advanced under fluoroscopy to a peripheral region of the lung, opened, closed and retracted obtaining a tissue sample; the hot forceps were employed in a similar fashion to the conventional forceps with the exception that, after grasping the tissue with the forceps, electrocautery was activated for approximately 5 s prior to pulling the sample from the surrounding lung tissue. The electrocautery generator (Olympus, Inc., Center Valley, Pa., USA) was set to 40 W for each biopsy. Samples were placed in formalin and processed for histology per standard hospital protocol. Bleeding was assessed as none, mild, moderate or severe after each pass of the biopsy forceps. This was adopted from prior clinical studies and defines bleeding as follows: no bleeding = presence of only traces of blood after finishing the biopsies with no need for continued suctioning; mild bleeding = need for continued suctioning of blood from the airways after the procedure; moderate bleeding = need for intubation of the biopsied segment with the flexible bronchoscope into the wedge position; severe bleeding = need for additional interventions, such as placing a temporary bronchus blocker, or intubation of the main stem bronchus on the bleeding side [11].

goto top of outline Histology

All samples were cut in 6-μm sections and stained with hematoxylin and eosin. A blinded pathologist (T.A.S.) examined all samples and measured each biopsy in the longest dimension. The areas of the biopsies containing alveolar tissue were identified and measured with an ocular micrometer, deriving an overall area of alveolar tissue (in mm2) for each biopsy. Examination of specimens at scanning magnification indicated a mean alveolar content of 61 alveoli/mm2. Alveoli were defined as sacs or alveoli proper bounded on at least 3 sides by septa. The alveolar content of each specimen was then determined as a function of mean alveolar content for the specimens in aggregate × area of alveolated tissue per biopsy. Each sample was assessed for artifact related to electrocautery application, which was noted as present or absent.

goto top of outline Statistics

All values are reported as means ± SD. The two-tailed t test was performed for comparisons between groups using the Statistics Toolbox for Matlab version 7.2 service pack 5 (Mathworks, Natick, Mass., USA). Differences were considered to be significant at a p value of <0.05.

 

goto top of outline Results

A total of 74 biopsies were collected using the different forceps types. Alveolar tissue was present in 25/74 and 26/74 of the biopsies obtained using the hot and conventional forceps, respectively. There was no difference in the size of biopsies collected (2.10 ± 1.10 vs. 1.83 ± 0.94 mm; p = 0.164) or the amount of alveoli per sample (343.2 ± 402.4 vs. 439.5 ± 463.5 alveoli; p = 0.433) for electrocautery and conventional forceps, respectively. There was no artifact related to the use of electrocautery, and bleeding was minimal using either forceps type.

 

goto top of outline Discussion

TBLB is a safe and effective method of diagnosing a variety of pulmonary conditions including infections, selected interstitial lung diseases, malignant neoplasms and lung transplant rejection [12]. Larger sample size is thought to have a greater diagnostic yield, although the use of larger forceps intuitively should lead to higher rates of significant bleeding and pneumothorax [3]. The ability to safely obtain a large lung tissue sample via TBLB would be a major advancement in the diagnosis of pulmonary diseases, particularly in interstitial lung disease, and would obviate the need to subject patients to surgical lung biopsy. A study by Babiak et al. [13] demonstrated the ability to obtain large lung tissue samples via TBLB using a cryoprobe with minimal side effects. Our study attempted to reach the same goal by using electrocautery-enabled large forceps to improve lung tissue sample size without increasing the risk.

We were unable to demonstrate a difference in biopsy size or alveolar content related to the use of the hot forceps. There are several possibilities as to why there was no discernable difference between the 2 forceps types: (1) the sampling area was limited to a specific region of the lung which may have resulted in anatomic structures limiting the opening of the forceps cups and restricting the sampling size; (2) swine, although thought to be a good model of human airways, may have differences in pulmonary anatomy which make them less amenable to TBLB, which is demonstrated by the low percentage of TBLB samples that had alveolar tissue, or (3) the variability in sample size is greater than any small improvement in sample size related to larger forceps, and a much larger study is needed to show the difference.

Significant bleeding is thought to occur in 1.9–8% of transbronchial biopsies due to the injury of a pulmonary arteriole by the biopsy forceps [14,15]. The theoretical advantage of the hot forceps is the reduction in bleeding complications related to the use of the electrocautery at the time of tissue sampling. In fact, a recent study using hot forceps for endobronchial biopsy of malignant lesions demonstrated a reduced, although clinically insignificant, reduction in bleeding [4]. We had no significant bleeding associated with the use of either forceps type. It is noteworthy that swine may not represent a good model for investigating bleeding related to TBLB. Previous studies investigating the association of bleeding with the use of clopidogrel and TBLB in swine showed no increased risk of bleeding, while subsequent studies in humans have established significantly more bleeding in patients taking clopidogrel at the time of TBLB [10,11].

Although our study findings are negative, we believe that it is important to share these results with the scientific community in order to advance the research of TBLB.

In conclusion, we have compared the use of conventional pulmonary forceps with larger electrocautery-enabled hot forceps for TBLB in swine. We found no improvement in sample size or size of alveolar tissue per sample when using the hot forceps. Further studies are needed to verify this finding in humans.

 

goto top of outline Acknowledgment

This study was supported by an educational grant from Olympus, Inc., Center Valley, Pa., USA.


 goto top of outline References
  1. Poletti V, Chilosi M, Olivieri D: Diagnostic invasive procedures in diffuse infiltrative lung diseases. Respiration 2004;71:107–119.
  2. Aleva RM, Kraan J, Smith M, et al: Techniques in human airway inflammation: quantity and morphology of bronchial biopsy specimens taken by forceps of three sizes. Chest 1998;113:182–185.
  3. Loube DI, Johnson JE, Wiener D, et al: The effect of forceps size on the adequacy of specimens obtained by transbronchial biopsy. Am Rev Respir Dis 1993;148:1411–1413.
  4. Casoni GL, Gurioli C, Chhajed PN, et al: The value of transbronchial lung biopsy using jumbo forceps via rigid bronchoscope in diffuse lung disease. Monaldi Arch Chest Dis 2008;69:59–64.
  5. Tremblay A, Michaud G, Urbanski SJ: Hot biopsy forceps in the diagnosis of endobronchial lesions. Eur Respir J 2007;29:108–111.
  6. Ram B, Oluwole M, Blair RL, et al: Surgical simulation: an animal tissue model for training in therapeutic and diagnostic bronchoscopy. J Laryngol Otol 1999;113:149–151.
  7. Brickey DA, Lawlor DP: Transbronchial biopsy in the presence of profound elevation of the international normalized ratio. Chest 1999;115:1667–1671.
  8. Hautmann H, Huber RM: Laser resistance of expandable metal stents in interventional bronchoscopy: an experimental evaluation. Lasers Surg Med 2001;29:70–72.
  9. Marquette CH, Mensier E, Copin MC, et al: Experimental models of tracheobronchial stenoses: a useful tool for evaluating airway stents. Ann Thorac Surg 1995;60:651–656.
  10. Wahidi MM, Garland R, Feller-Kopman D, et al: Effect of clopidogrel with and without aspirin on bleeding following transbronchial lung biopsy. Chest 2005;127:961–964.
  11. Ernst A, Eberhardt R, Wahidi M, et al: Effect of routine clopidogrel use on bleeding complications after transbronchial biopsy in humans. Chest 2006;129:734–737.
  12. Fulkerson WJ: Current concepts. Fiberoptic bronchoscopy. N Engl J Med 1984;311:511–515.
  13. Babiak A, Hetzel J, Krishna G, et al: Transbronchial cryobiopsy: a new tool for lung biopsies. Respiration 2009;78:203–208.
  14. Bjortuft O, Brosstad F, Boe J: Bronchoscopy with transbronchial biopsies: measurement of bleeding volume and evaluation of the predictive value of coagulation tests. Eur Respir J 1998;12:1025–1027.
  15. Cordasco EM Jr, Mehta AC, Ahmad M: Bronchoscopically induced bleeding. A summary of nine years’ Cleveland clinic experience and review of the literature. Chest 1991;100:1141–1147.

 goto top of outline Author Contacts

Momen M. Wahidi, MD, MBA
Interventional Pulmonology, Division of Pulmonary and Critical Care Medicine
Duke University Medical Center, Box 3683
Durham, NC 27710 (USA)
Tel. +1 919 668 0340, Fax +1 919 684 3874, E-Mail momen.wahidi@duke.edu


 goto top of outline Article Information

Received: March 17, 2009
Accepted after revision: June 19, 2009
Published online: August 25, 2009
Number of Print Pages : 4
Number of Figures : 2, Number of Tables : 0, Number of References : 15


 goto top of outline Publication Details

Respiration (International Journal of Thoracic Medicine)

Vol. 79, No. 2, Year 2010 (Cover Date: December 2009)

Journal Editor: Bolliger C.T. (Cape Town)
ISSN: 0025-7931 (Print), eISSN: 1423-0356 (Online)

For additional information: http://www.karger.com/RES


Copyright / Drug Dosage / Disclaimer

Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher or, in the case of photocopying, direct payment of a specified fee to the Copyright Clearance Center.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in goverment regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.

Abstract

Background: Transbronchial lung biopsy (TBLB) is a commonly performed bronchoscopic procedure. Previous studies have suggested that larger biopsy forceps may improve diagnostic yield; however, the risk of bleeding associated with larger samples may be increased. The hot forceps are large forceps that are connected to an electrocautery system to minimize bleeding at the time of biopsy. Objectives: We evaluated the hot forceps for improvement in biopsy size and the number of sampled alveoli. Methods: TBLBs were performed in 2 swine using one type of the forceps, followed by the other forceps 24 h later. Electrocautery was applied from closure of the forceps to retrieval of the sample. A blinded pathologist measured the size of each sample in its longest dimension and calculated the total alveolar content within the largest cross-section from each biopsy. Results: A total of 74 biopsies were collected using each forceps type. Alveolar tissue was present in 25/74 and 26/74 of the biopsies using the hot and conventional forceps, respectively. There was no difference in the size of biopsies collected (2.10 ± 1.10 vs. 1.83 ± 0.94 mm; p = 0.164) or in the amount of alveoli per sample (343.2 ± 402.4 vs. 439.5 ± 463.5 alveoli; p = 0.433) for hot and conventional forceps, respectively. There was no artifact related to the use of electrocautery, and bleeding was minimal using either forceps system. Conclusions: The use of the electrocautery hot forceps for TBLB did not result in improvement of the size of biopsies or the amount of collected alveolar tissue in healthy pigs.



 goto top of outline Author Contacts

Momen M. Wahidi, MD, MBA
Interventional Pulmonology, Division of Pulmonary and Critical Care Medicine
Duke University Medical Center, Box 3683
Durham, NC 27710 (USA)
Tel. +1 919 668 0340, Fax +1 919 684 3874, E-Mail momen.wahidi@duke.edu


 goto top of outline Article Information

Received: March 17, 2009
Accepted after revision: June 19, 2009
Published online: August 25, 2009
Number of Print Pages : 4
Number of Figures : 2, Number of Tables : 0, Number of References : 15


 goto top of outline Publication Details

Respiration (International Journal of Thoracic Medicine)

Vol. 79, No. 2, Year 2010 (Cover Date: December 2009)

Journal Editor: Bolliger C.T. (Cape Town)
ISSN: 0025-7931 (Print), eISSN: 1423-0356 (Online)

For additional information: http://www.karger.com/RES


Copyright / Drug Dosage

Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher or, in the case of photocopying, direct payment of a specified fee to the Copyright Clearance Center.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in goverment regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.

References

  1. Poletti V, Chilosi M, Olivieri D: Diagnostic invasive procedures in diffuse infiltrative lung diseases. Respiration 2004;71:107–119.
  2. Aleva RM, Kraan J, Smith M, et al: Techniques in human airway inflammation: quantity and morphology of bronchial biopsy specimens taken by forceps of three sizes. Chest 1998;113:182–185.
  3. Loube DI, Johnson JE, Wiener D, et al: The effect of forceps size on the adequacy of specimens obtained by transbronchial biopsy. Am Rev Respir Dis 1993;148:1411–1413.
  4. Casoni GL, Gurioli C, Chhajed PN, et al: The value of transbronchial lung biopsy using jumbo forceps via rigid bronchoscope in diffuse lung disease. Monaldi Arch Chest Dis 2008;69:59–64.
  5. Tremblay A, Michaud G, Urbanski SJ: Hot biopsy forceps in the diagnosis of endobronchial lesions. Eur Respir J 2007;29:108–111.
  6. Ram B, Oluwole M, Blair RL, et al: Surgical simulation: an animal tissue model for training in therapeutic and diagnostic bronchoscopy. J Laryngol Otol 1999;113:149–151.
  7. Brickey DA, Lawlor DP: Transbronchial biopsy in the presence of profound elevation of the international normalized ratio. Chest 1999;115:1667–1671.
  8. Hautmann H, Huber RM: Laser resistance of expandable metal stents in interventional bronchoscopy: an experimental evaluation. Lasers Surg Med 2001;29:70–72.
  9. Marquette CH, Mensier E, Copin MC, et al: Experimental models of tracheobronchial stenoses: a useful tool for evaluating airway stents. Ann Thorac Surg 1995;60:651–656.
  10. Wahidi MM, Garland R, Feller-Kopman D, et al: Effect of clopidogrel with and without aspirin on bleeding following transbronchial lung biopsy. Chest 2005;127:961–964.
  11. Ernst A, Eberhardt R, Wahidi M, et al: Effect of routine clopidogrel use on bleeding complications after transbronchial biopsy in humans. Chest 2006;129:734–737.
  12. Fulkerson WJ: Current concepts. Fiberoptic bronchoscopy. N Engl J Med 1984;311:511–515.
  13. Babiak A, Hetzel J, Krishna G, et al: Transbronchial cryobiopsy: a new tool for lung biopsies. Respiration 2009;78:203–208.
  14. Bjortuft O, Brosstad F, Boe J: Bronchoscopy with transbronchial biopsies: measurement of bleeding volume and evaluation of the predictive value of coagulation tests. Eur Respir J 1998;12:1025–1027.
  15. Cordasco EM Jr, Mehta AC, Ahmad M: Bronchoscopically induced bleeding. A summary of nine years’ Cleveland clinic experience and review of the literature. Chest 1991;100:1141–1147.