Kidney and Blood Pressure Research
Original Paper
Wrist Measurement of Blood Pressure: Some Critical Remarks to OscillometryWeber F. · Erbel R. · Schäfers R. · Philipp T.Zentrum für Innere Medizin, Universitätsklinik Essen, Deutschland
|
|
Log in to MyKarger to check if you already have access to this content.
KAB
Buy a Karger Article Bundle (KAB) and profit from a discount!
If you would like to redeem your KAB credit, please log in.
Save over 20% compared to the individual article price.
Article / Publication Details
Published online: June 25, 1999
Issue release date: 1999
Number of Print Pages: 5
Number of Figures: 4
Number of Tables: 0
ISSN: 1420-4096 (Print)
eISSN: 1423-0143 (Online)
For additional information: https://www.karger.com/KBR
Abstract
Oscillometric blood pressure measuring devices worn at the wrist are preferred by a growing part of hypertensive patients for self–measurement because of their compactness and ease of use. However, little is known about the accuracy of such instruments. The accuracy of such devices was evaluated by measuring the blood pressure at the end of a coronary angiography procedure in 27 subjects nearly simultaneously in the aortic arch, using a Statham P23 transducer, and at the wrist, using the blood pressure watch (BPW) by NAIS–Matsushita. Four replicate comparative measurements were performed in each subject. On average, the blood pressure measured by the BPW was found to be higher: systolic +1.2±(SD)10.2 mm Hg, diastolic +4.1±7.2 mm Hg. The correlation coefficient between the two methods was 0.85 for the systolic and 0.84 for the diastolic pressure, but 39% of the systolic and 22% of the diastolic differences were outside the range of ±1 SD. There was no correlation between systolic differences and blood pressure or patients’ variables, but multiple regression analysis revealed that the diastolic blood pressure differences were negatively correlated with the aortic diastolic pressure (r = 0.89, p<0.001), indicating that the BPW is less accurate to detect lower values. Conclusions: (1) Though the average systolic and diastolic deviations between both methods and the relative correlation coefficients suggest that the BPW is to be used carefully for monitoring blood pressure, the great variability of differences and the inability of the BPW to detect lower diastolic blood pressure calls for a mandatory validation test by an approved protocol before this new device can be recommended for general use. (2) On the basis of our results for wrist oscillometry with the BPW and data reported in the literature for upper–arm oscillometric devices, we recommend to use oscillometric devices only when they were validated properly. In particular, the oscillometric measurement should be employed with caution for epidemiological studies or for investigation of population groups (e.g., children or pregnant women) that are expected to display lower diastolic pressures.
Related Articles:
References
- Audet, A–M for the American College of Physicians: Automated ambulatory blood pressure and self–measured blood pressure monitoring devices: Their role in the diagnosis and management of hypertension. Ann Intern Med 1993;118:889–892.
- Pickering TG: Blood pressure measurement and detection of hypertension. Lancet 1994; 344:31–35.
-
Anlauf M, Tholl U, Högermeyer D, Hirche H, Roggenbuck U, Simonides R: Devices for blood pressure self–measurement: Tested in comparison. Z Kardiol 1996;85(suppl 3):20– 25.
-
Weber F, Hirche H, Simonides R, Anlauf M: Comparison of the accuracy of electronic devices for the self–measurement of blood pressure produced between 1980 and 1983. Z Kardiol 1985;74:618–623.
-
Kribben A, Weber F, Anlauf M, Claus M, Hirche H, Simonides R, Philipp T: Electronic devices for the self–measurement of blood pressure: Comparison of oscillometric and Korotkoff devices. Nieren Hochdruckkr 1992;21: 351–357.
-
Association for the Advancement of Medical Instrumentation: American National Standard for Electronic or Automated Sphygmomanometers. Arlington, AAMI, 1993.
-
O’Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman DG, Bland JM, Coats A, Atkins N: The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993;11(suppl 2):43–62.
- Mancia G, Parati G: Commentary on the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices: A critique of aspects related to 24–hour ambulatory blood pressure measurement. J Hypertens 1993;11:595–597.
- Gardner RM: Direct blood pressure measurement–dynamic response requirements. Anesthesiology 1981;54:227–236.
-
Altman DG, Bland JM: Analysis of blood pressure data; Birkenhäger WH, Reid JL, (eds): Handbook of Hypertension. Amsterdam, Elsevier, 1991, vol 14, pp 287–314.
-
Eckert S, Gleichmann S, Gleichmann U: Blood pressure measurement in the wrist is accurate and reliable (abstract). 18th Meet of the German Hypertension League, Joint Meeting with the Dutch Hypertension Society, Münster 1994.
-
Klaus D, Wetzchewald D, Saul F, Obermeier G, Lambers F: Comparison of blood pressure measurements by semi–automatic cuff–oscillometric and sphygmomanometric microphone methods with intra–arterial values. Nieren Hochdruckkr 1991;20:160–169.
- Pannarale G, Bebb G, Clark S, Sullivan A, Foster C, Coates AJS: Bias and variability in blood pressure measurement with ambulatory recorders. Hypertension 1993;22:591–598.
- Lurbe E, Redon J, Liao Y, Tacons J, Cooper RS, Alvarez V: Ambulatory blood pressure monitoring in normotensive children. J Hypertens 1994;12:1417–1423.
- Perloff D, Sokolow M, Cowan R: The prognostic value of ambulatory blood pressure. JAMA 1983;249:2792–2798.
- Baumgart P, Walger P, Jürgens U, Rahn KH: Reference data for ambulatory blood pressure monitoring: What results are equivalent to the established limits of office blood pressure? Klin Wochenschr 1990;68:723–727.
Article / Publication Details
Published online: June 25, 1999
Issue release date: 1999
Number of Print Pages: 5
Number of Figures: 4
Number of Tables: 0
ISSN: 1420-4096 (Print)
eISSN: 1423-0143 (Online)
For additional information: https://www.karger.com/KBR
Copyright / Drug Dosage / Disclaimer
Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher.Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.

Get Permission