Technical Note

Free Access

Agreement between Spirometers: A Challenge in the Follow-Up of Patients and Populations?

Gerbase M.W.a · Dupuis-Lozeron E.a, b · Schindler C.c, d · Keidel D.c · Bridevaux P.O.a · Kriemler S.c, d · Probst-Hensch N.M.c, d · Rochat T.a · Künzli N.c, d

Author affiliations

aDivision of Pulmonary Medicine, University Hospitals of Geneva, and bResearch Center for Statistics and Department of Management Sciences, University of Geneva, Geneva, and cSwiss Tropical and Public Health Institute and dUniversity of Basel, Basel, Switzerland

Corresponding Author

Margaret W. Gerbase, MD, PhD

Division of Pulmonary Medicine, University Hospitals of Geneva

4 rue Gabrielle-Perret-Gentil, CH-1211 Geneva 14 (Switzerland)

E-Mail margaret.gerbase@hcuge.ch

Related Articles for ""

Respiration 2013;85:505-514

Do you have an account?

Login Information





Contact Information












By signing up for MyKarger you will automatically participate in our year-End raffle.
If you Then Do Not wish To participate, please uncheck the following box.

Yes, I wish To participate In the year-End raffle And Get the chance To win some Of our most interesting books, And other attractive prizes.


I have read the Karger Terms and Conditions and agree.



Abstract

Background: Long-term cohort studies and lung function laboratories are confronted with the need for replacement of spirometers. Lack of agreement between spirometers might affect the longitudinal comparison of data, notably when replacing conventional by portable spirometers. Objectives: To compare the handheld EasyOne (EO) with the conventional SensorMedics (SM) spirometer, and to analyze the interdevice reproducibility of EO spirometers. Methods: In total, 82 volunteers completed spirometry sessions with 1 SM and 2 of 3 EO spirometers following a Latin square design. Analyses of differences in forced vital capacity (FVC), forced expiratory flow in 1 s (FEV1), FEV1/FVC and mean forced expiratory flow calculated between 25 and 75% of the FVC between spirometers used a mixed effect model with a random intercept for each subject and the effect of the device as fixed effect adjusted for sex, age, height and order of spirometer tested. Bland-Altman plots show the 95% limits of agreement. Results: Comparisons between EO and SM showed relatively small mean differences of <3%, but systematically lower values for FVC and FEV1 in all EO devices. The 95% agreement exceeded the limits for FEV1 by 50 ml in 2 EO spirometers. The EO interdevice comparisons showed mean differences and limits of agreement within established thresholds, thus indicating fair accuracy when comparing devices. Repeats with the same spirometer did not result in statistically significant differences. Conclusions: This study suggests fair agreement between the handheld and the conventional spirometer. Differences slightly exceeding limits for FEV1 in 2 EO devices might be considered mostly irrelevant for clinical practice. However, the systematically lower FVC and FEV1 observed with EO may be significant for epidemiological studies, thus justifying inspection before replacing devices.

© 2013 S. Karger AG, Basel


References

  1. Caras WE, Winter MG, Dillard T, Reasor T: Performance comparison of the handheld MicroPlus portable spirometer and the SensorMedics Vmax22 diagnostic spirometer. Respir Care 1999;44:1465-1473.
  2. Viegi G, Simoni M, Pistelli F, Englert N, Salonen R, Niepsuj G, et al: Inter-laboratory comparison of flow-volume curve measurements as quality control procedure in the framework of an international epidemiological study (PEACE project). Respir Med 2000;94:194-203.
  3. Swart F, Schuurmans MM, Heydenreich JC, Pieper CH, Bolliger CT: Comparison of a new desktop spirometer (Spirospec) with a laboratory spirometer in a respiratory out-patient clinic. Respir Care 2003;48:591-595.
    External Resources
  4. Korhonen H, Remes ST, Kannisto S, Korppi M: Hand-held turbine spirometer: agreement with the conventional spirometer at baseline and exercise. Pediatr Allergy Immunol 2005;16:254-257.
  5. Akhtar R, Wilson A: A comparison of spirometry in general practice and a pulmonary function laboratory. Prim Care Respir J 2005;14:215-220.
  6. Liistro G, Vanwelde C, Vincken W, Vandevoorde J, Verleden G, Buffels J, et al: Technical and functional assessment of 10 office spirometers: a multicenter comparative study. Chest 2006;130:657-665.
  7. Walters JAE, Wood-Baker R, Walls J, Johns DP: Stability of the EasyOne ultrasonic spirometer for use in general practice. Respirology 2006;11:306-310.
  8. Skloot GS, Edwards NT, Enriight PL: Four-year calibration stability of the EasyOne portable spirometer. Respir Care 2010;55:873-877.
    External Resources
  9. Perez-Padilla R, Vasquez-Garcia JC, Marquez MN, Menezes AMB, PLATINO Group: Spirometry quality-control strategies in a multinational study of the prevalence of chronic obstructive pulmonary disease. Respir Care 2008;53:1019-1026.
    External Resources
  10. Enright P, Vollmer WM, Lamprecht B, Jensen R, Jithoo A, Tan W, et al: Quality of spirometry tests performed by 9893 adults in 14 countries: the BOLD study. Respir Med 1995;105:1507-1515.
  11. Barr RG, Stemple KJ, Mesia-Vela S, Basner RC, Derk SJ, Hennenberger PK, et al: Reproducibility and validity of a handheld spirometer. Respir Care 2008;53:433-441.
    External Resources
  12. Martin BW, Ackermann-Liebrich U, Leuenberger P, Künzli N, Stutz EZ, Keller R, et al: SAPALDIA: methods and participation in the cross-sectional part of the Swiss Study on Air Pollution and Lung Diseases in Adults. Soz Präventivmed 1997;42:67-84.
  13. Ackermann-Liebrich U, Kuna-Dibbert B, Probst-Hensch NM, Schindler C, Felber Dietrich D, Stutz EZ, et al: Follow-up of the Swiss Cohort Study on Air Pollution and Lung Diseases in Adults (SAPALDIA 2) 1991-2003: methods and characterization of participants. Soz Präventivmed 2005;50:245-263.
  14. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, et al: Standardisation of spirometry. Eur Respir J 2005;26:319-338.
  15. Standardization of spirometry, 1994 update. American Thoracic Society. Am J Respir Crit Care Med 1995;152:1107-1136.
  16. Standardization of spirometry - 1987 update. Statement of the American Thoracic Society. Am Rev Respir Dis 1987;136:1285-1298.
    External Resources
  17. Brändli O, Schindler C, Künzli N, Keller R, Perruchoud AP: Lung function in healthy never smoking adults: reference values and lower limits of normal of a Swiss population. Thorax 1996;51:277-283.
  18. Pinheiro JC, Bates DM: Mixed-Effects Models in S and S-PLUS. Statistics and Computing Series. New York, Springer, 2000.
  19. Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-310.
    External Resources
  20. R Development Core Team: R: A Language and Environment for Statistical Computing. Vienna, R Foundation for Statistical Computing, 2012.
  21. Pinheiro J, Bates D, DebRoy S, Sarkar D, R Development Core Team: NLME: linear and nonlinear mixed effects models. R package version 3.1-104, 2012. R-core@R-project.org.
  22. Carstensen B, Gurrin L, Ekstrom C: MethComp: functions for analysis of method comparison studies. R package version 1.15, 2012. http://CRAN.R-project.org/package=MethComp.
  23. Wickham H: ggplot2: Elegant Graphics for Data Analysis. New York, Springer, 2009.
  24. Schoh RJ, Fero LJ, Shapiro H, Aslor JP, Kaelin OJ, Rollins DR, et al: Performance of a new screening spirometer at a community fair. Respir Care 2002;47:1150-1157.
    External Resources
  25. Perez-Padilla R, Vasquez-Garcia JC, Marquez MN, Jardim JRB, Pertuzé J, Lisboa C, et al: The long-term stability of portable spirometers used in a multinational study of the prevalence of chronic obstructive pulmonary disease. Respir Care 2006;51:1167-1171.
    External Resources
  26. Chinn S, Jarvis D, Svanes C, Burney P: Sources of variation in forced expiratory volume in one second and forced vital capacity. Eur Respir J 2006;27:767-773.
  27. Künzli N, Ackermann-Liebrich U, Keller R, Perruchoud AP, Schindler C: Variability of FVC and FEV1 due to technician, team, device and subject in an eight centre study: three quality control studies in SAPALDIA. Swiss Study on Air Pollution and Lung Disease in Adults. Eur Respir J 1995;8:371-376.
  28. Künzli N, Kuna-Dibbert B, Keidel D, Keller R, Brändli O, Schindler C, et al: Longitudinal validity of spirometers - a challenge in longitudinal studies. Swiss Med Wkly 2005;135:503-508.
  29. Schermer TRJ, Verweij EHA, Cretier R, Pellegrino A, Crockett AJ, Poels PJP: Accuracy and precision of desktop spirometers in general practices. Respiration 2012;83:344-352.

Article / Publication Details

First-Page Preview
Abstract of Technical Note

Received: October 11, 2012
Accepted: December 13, 2012
Published online: March 08, 2013
Issue release date: May 2013

Number of Print Pages: 10
Number of Figures: 2
Number of Tables: 4

ISSN: 0025-7931 (Print)
eISSN: 1423-0356 (Online)

For additional information: https://www.karger.com/RES


Copyright / Drug Dosage / Disclaimer

Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.